Differential pressure measurement is widely used as a way for monitoring the level of closed vessels. In many cases, operators use individual instruments with an output signal or on-site display. On the other hand, a far more precise and fault-immune solution may be the measuring arrangement developed by WIKA using two process transmitters in a primary-secondary relationship.
เครื่องมือวัดความดัน monitoring application using differential pressure measurement is founded on the next principle: For the amount of the contents in the vessel, the differential pressure between your liquid and the gaseous phase is set. In addition, the hydrostatic pressure, the precise density of the medium and the vessel geometry are included in the calculation. Typical measuring instruments have two adjacent process connections for the pressures P1 and P2. For level measurement, a capillary must bridge the pronounced distance between the two measuring points.
Measurement with two process transmitters as Primary instrument (right) and Secondary instrument. This measuring arrangement for level monitoring may be used flexibly, e.g. for home elevators the mixing ratio of liquids.
Connection via signal cable
When monitoring the particular level using electronic differential pressure measurement, however, both measuring points have one process transmitter each. For instance, the WIKA models CPT-2x and IPT-2x are suitable for this. The two transmitters are arranged as Primary instrument and Secondary instrument. They are connected, purely electrically, via a signal cable and so are therefore not vunerable to disturbances. The Primary instrument supplies the Secondary instrument with power. The Secondary instrument is parameterised via a serial interface or through the display on the Primary instrument. Finally, communication takes place via an interior bus.
Level monitoring in a closed tank
Application example
The figure on the left shows a typical application example for level monitoring with electronic differential pressure measurement on a liquid tank. The Secondary instrument measures the pressure P1, in cases like this the gas pressure, and transmits it to the Primary instrument. This detects the pressure P2, here the pressure in the liquid column at the bottom of the tank. THE PRINCIPAL instrument calculates the differential pressure from P1 and P2. Using the measured value and considering the other parameters mentioned, it determines the volume in the tank as an indication of the level. This value could be output either being an analogue or digital signal to the control room and/or to the display of the principal instrument.
Benefits of the measuring arrangement with two process transmitters
The technique for level monitoring described here has several advantages when compared to a conventional method, for instance, with a differential pressure transmitter:
Higher accuracy and ?turndown?
Process transmitters deliver higher levels of accuracy. In the case of the WIKA models, these are around 0.05 % of the set span. In addition, there is the chance of ?turndown?, the average person setting of a specific measuring span (that is also possible retrospectively with instruments that are already installed.)
Minimised temperature effect
Transmission via signal cable with electronic differential pressure measurement minimises the temperature effect. In contrast, the capillaries found in the conventional measuring method are much more susceptible here. This, in turn, can ultimately affect the measuring result.
Fast commissioning
The electronic measuring arrangement could be commissioned faster. Contrary to a remedy with capillaries, it does not require any test run.
Reduced maintenance effort
The maintenance effort is leaner: In case of a fault, only the process transmitter concerned needs to be replaced. On the other hand, with measuring differential pressure with capillaries, the entire measuring arrangement should be replaced.
Note
Further information on the model CPT-2x and IPT-2x process transmitters, which are ideal for electronic differential pressure measurement for level monitoring, is available on the WIKA website.

Leave a Reply

Your email address will not be published. Required fields are marked *